Images are representations only.

3300HT/3300HTVP PERpH-X High Performance pH/ORP Sensor

Brand: Rosemount
Rosemount 3300HT and 3300HTVP PERpH-X High Performance pH/ORP sensors feature a robust design that significantly prolongs sensor life in complex applications. 

These rebuildable sensors work well in high temperature and high pressure processes and resist fouling, coating, and poisoning. 

Compared to other sensor designs, the 3300HT and 3300HTVP decrease the frequency of sensor replacement and provides you with faster pH/ORP response times and decreased drift.
Features
  • Increased sensor life and performance in high temperature and high pressure applications up to 155° C and 400 psig
  • SMART pH compatibility with Rosemount pH transmitters
  • A rebuildable design with various electrolytes to best suit your application needs
  • Ease of installation with Variopol (VP8) cable connection
Rosemount 3300HT and 3300HTVP PERpH-X High Performance pH/ORP sensors feature a robust design that significantly prolongs sensor life in complex applications. 

These rebuildable sensors work well in high temperature and high pressure processes and resist fouling, coating, and poisoning. 

Compared to other sensor designs, the 3300HT and 3300HTVP decrease the frequency of sensor replacement and provides you with faster pH/ORP response times and decreased drift.
Features
  • Increased sensor life and performance in high temperature and high pressure applications up to 155° C and 400 psig
  • SMART pH compatibility with Rosemount pH transmitters
  • A rebuildable design with various electrolytes to best suit your application needs
  • Ease of installation with Variopol (VP8) cable connection

Specifications

Approvals & Certifications

  • ATEX
  • CSA
  • FM
  • IECEx

Materials of Construction

  • Electrode: Glass

Maximum System Pressure

  • psia: 414.66 (kPa: 2,859, bar: 28.59)
  • psig: 400 (kPa: 2,758, bar: 27.58) (2,859 kPa [abs])

Measuring Range

  • 0 to 14 pH
  • ORP: -1,500 to 1,500 mV

O-Ring Material

  • EPDM
  • Kalrez
  • Viton

Operating Temperature

  • 5° to 155° C (41° to 311° F) without Preamplifier
  • Up to 100° C (212° F) with Preamplifier

Pressure Rating

  • CRN Rating - psig: 200 (kPa: 1,379, bar: 13.79) at Room Temperature

RTD Type

  • Pt100

Storage Temperature

  • -10° to 70° C (14° to 158° F)

Wetted Materials

  • Glass
  • Ryton
  • Teflon
  • Titanium

Documentation